A pump error can occur when the temperature and pressure changes, there is a leak or block in the pneumatics or if the calibration points are incorrect. Navigate to the main menu and select Calibration Management. Within this menu is an option to perform a pump calibration. If the pump calibration does not fix the pump error, then load and new calibration file. If there is still a pump error please contact an ION Science Authorised Service Centre.